采用湿式消解法对金线莲药材进行处理,石墨炉原子吸收光谱法测定其铅(Pb)、铜(Cu)、镉(Cd)等重金属元素含量。结果表明,Pb、Cu、Cd 三种元素的原子化温度分别是2100、2200和1800 ℃,灰化温度分别是400、800和250 ℃。Pb、Cu、Cd三种元素的检出限分别为0.47、0.50、0.62 μg·L-1,其加标回收率在92.7%~94.0%之间,平均相对标准偏差为1.71%,说明仪器精密度良好,湿式消解-石墨炉原子吸收光谱法能有效测定金线莲药材中Pb、Cu、Cd含量。
The samples were treated by wet digestion method to measure the content of Pb, Cu, Cd in Anoectochilus roxburghii by using graphite furnace atomic absorption spectrometry (GFAAS). The results showed that the atomization temperature of Pb, Cu, Cd were 2100, 2200 and 1800 ℃, the ashing temperature were 400, 800 and 250 ℃, and their detection limits were 0.47, 0.50 and 0.62 μg·L-1, respectively. The spike recoveries ranged from 92.7% to 94.0%, and the average precision relative standard deviation was 1.71%, which showed that this method is effective and reliable for determination of Pb, Cu, Cd in A. roxburghii.
[1]《全国中草药汇编》编写组. 全国中草药汇编(下册)[M]. 北京: 人民卫生出版社, 1992.
[2] 唐明娟,孟志霞,郭顺星. 菌根真菌对人工栽培金线莲糖类和无机元素的影响[J]. 中草药, 2008(10): 1565—1568.
[3] 陈晋红,汤毅珊,刘大伟,王宁生. 姜黄药材中6种重金属残留量测定[J]. 中药新药与临床药理, 2009(5): 457—460.
[4] 中华人民共和国卫生部, 中国国家标准化管理委员会. GB/T5009.12—2003食品中铅的测定[S]. 北京: 中国标准出版社, 2003.
[5] 中华人民共和国卫生部, 中国国家标准化管理委员会. GB/T5009.13—2003食品中铜的测定[S]. 北京: 中国标准出版社, 2003.
[6] 中华人民共和国卫生部, 中国国家标准化管理委员会. GB/T5009.15—2003食品中镉的测定[S]. 北京: 中国标准出版社, 2003.
[7] 胡国海,李洪潮,解成骏. 云南文山人工种植金线莲中的微量元素含量测定[J]. 安徽农业科学, 2010,38(14): 7294—7295, 7330.
[8] 李洪潮,胡国海,解成骏. 云南文山野生金线莲中微量元素研究[J]. 食品工业科技, 2010(10): 386—388.
[9] 刘童,王永林,王爱民,兰燕宇,李勇军. 微波消解-石墨炉原子吸收法测定荭草药材中的铅、镉、铜含量[J]. 贵阳医学院学报, 2010,35(6): 551—554.
[10] 吴彩霞,张洪荣. 微波消解石墨炉原子吸收光谱法测定植物中的铅[J]. 草业科学, 2010,27(2): 66—70.
[11] 徐小艳,孙远明,苏文焯,谢郁文. 微波消解-石墨炉原子吸收光谱法连续测定水果和蔬菜中铅铬镉[J]. 食品科学, 2009,30(10): 206—208.